Discuss your analysis

PARENTERAL / SUBVISIBLE PARTICLE STUDIES

Subvisible Particle Characterization

Subvisible Particle Characterization

Understand particle populations in parenteral products and compare study-defined samples. Automated imaging provides counts, size distributions and morphology; Raman measurements of selected particles add chemical material information.

One study. Clearly defined measurement populations.

Optically detected and countable particles form the population measured under the agreed imaging and preparation conditions. Calibrated morphology describes those observed particles without assigning chemical identity.

Raman selection defines the targets chosen for spectral measurement. Report the actual Raman-measured subset separately from accepted chemical classifications, inconclusive spectra and unresolved assignments. An unmeasured particle is not an unresolved Raman measurement.

State the relevant denominator with every result. Selected Raman measurements do not by themselves validate extrapolation of material shares to every optically counted particle.

Measured evidence from an autoinjector investigation

The three original A02 modules below connect measured size-class counts, material-class comparison and particle-level Raman evidence. Their figures, captions, size classes, references and qualifications are retained unchanged.

The open-ended >25 µm bin is not a universal boundary of visibility. The material-class bubble figure illustrates relative representation within a batch; it is not a quantitative abundance scale or a formal equivalence test.

Compare particle size distributions across pharmaceutical samples

SizeID.bio generates its own documented particle counts and size distributions using calibrated imaging. Compare measured counts across size classes to examine differences between investigated samples.

Particle size distributions from ten pharmaceutical autoinjector study samples, comparing particle counts across size classes from 2 µm to above 25 µm.
Particle size distributions from ten pharmaceutical autoinjector study samples, comparing particle counts across size classes from 2 µm to above 25 µm.

Particle size distributions across ten representative samples from a pharmaceutical autoinjector investigation, showing measured particle counts in defined size classes.

Particle size distributions across ten representative samples from a pharmaceutical autoinjector investigation, showing measured particle counts in defined size classes.

Light obscuration and microscopy use different measurement principles; their counts are not automatically equivalent. The size classes in this example start at 2 µm. The achievable particle-size range and Raman identification performance depend on the material, sample matrix, preparation and study-specific analytical method.

View full-size figure

Which materials dominate the particle population?

Raman material-class comparison reveals compositional differences between samples. Chemical population fingerprints show which materials dominate and how the investigated populations differ.

Raman chemical particle population fingerprints comparing rubber, silicone, polyamide, polypropylene, polyethylene, cellulose and other material classes across ten study batches.
Raman chemical particle population fingerprints comparing rubber, silicone, polyamide, polypropylene, polyethylene, cellulose and other material classes across ten study batches.

Raman-identified material classes across ten investigated particle populations. Relative class representation visualizes differences in chemical composition between samples.

Raman-identified material classes across ten investigated particle populations. Relative class representation visualizes differences in chemical composition between samples.

Material classes derive from Raman measurements. Bubble sizes are a qualitative illustration of relative class representation within each batch, not a quantitative abundance scale or a formal statistical equivalence test.

Material classes derive from Raman measurements. Bubble sizes are a qualitative illustration of relative class representation within each batch, not a quantitative abundance scale or a formal statistical equivalence test.

View full-size figure

Identify the materials behind the particle counts

Original microscopy images and Raman spectra identify rubber, silicone, polyamide and epoxy-like materials. Comparison with suitable component references can suggest possible links, not proven root causes.

Raman-identified pharmaceutical device particles with microscopy images and spectra of rubber, silicone, polyamide and epoxy-like materials, illustrating potential contamination sources.
Raman-identified pharmaceutical device particles with microscopy images and spectra of rubber, silicone, polyamide and epoxy-like materials, illustrating potential contamination sources.

Representative particle images and Raman spectra for rubber, silicone, polyamide and epoxy-like materials, with potential links to pharmaceutical delivery-device components.

Representative particle images and Raman spectra for rubber, silicone, polyamide and epoxy-like materials, with potential links to pharmaceutical delivery-device components.

Microscopy images and Raman spectra are measured evidence. The device-component illustrations indicate possible source relationships; definitive source attribution requires suitable reference materials, sampling controls and investigation context.

Microscopy images and Raman spectra are measured evidence. The device-component illustrations indicate possible source relationships; definitive source attribution requires suitable reference materials, sampling controls and investigation context.

View full-size figure

Discuss a similar investigation →

Discuss a similar investigation →

Explore foreign particle source investigation →

Methods defined for the product and intended use

We agree the sample preparation, optical measurements, Raman-selection strategy and comparative endpoints for the study. Background, recovery, matrix effects and particle alteration or loss are considered before interpretation.

Explore Automated Particle Raman Analysis

Regulatory context — complementary analytical information

Applicable product requirements may include USP <788> and Ph. Eur. 2.9.19, with the relevant USP <1788> method guidance. Automated Particle Raman Analysis provides complementary morphological and chemical information; it is not automatically interchangeable with applicable compendial counting methods or a batch-release decision.

Therapeutic-protein and formulation context

Biological products need product-specific interpretation. USP <1787> and the applicable <787>/<788> framework provide relevant context; a proposed change to <787> is not assumed to be an already effective omission. The applicable edition and intended use are qualified when the detailed current text has not been verified.

USP <1787> — informational characterization context

USP <787> — therapeutic-protein chapter and transition context

Visible particle defects and investigations

Difficult-to-inspect products and preparation methods

Parenteral programme overview

Compare your particle populations

Tell us about the product, sample matrix and question. We can scope the measurements and evidence your investigation needs.

Discuss your project